Skip to content

HPLC preparative systems

PLC 2020 (GILSON)

The PLC 2020 is a compact and self-contained purification platform that can perform manual injection purification runs from 1 to 100 mL/min with pressures up to 4060 psi.

The system can accommodate up to four 1-liter solvent tanks and choose from four different elution solvents. It also features integrated dual wavelength variable UV/VIS detection and fraction collection functions, all controlled by the user-friendly touchscreen controller with an intuitive, easy-to-use software interface.

-Solvent selection (up to 4 solvents selectable on Pump B, 20 μm inlet solvent filter)

-Pumping System 100SC Pump Head (Flow: 1–100 mL/min,  Pressure: 1015 psi, 70 bars)

-Injection (Electronically-actuated manual injection valve, 5 mL loop installed)

-columns : XBridge peptide BEH C18 OBD prep column (5 mm, 300 Å, 10 × 250 mm), XBridge Prep C18 5 µm OBD, 19 × 150 mm, or 19 × 100 mm and Zorbax 300SB-C3 column (5 µm, 9.4 × 250 mm) are placed in an extra temperature chamber, which can be programed (up to 70 °C).

-Detector (Microprocessor-controlled absorbance detector providing dual variable UV/VIS wavelength detection, Range : 190–700 nm).

Pure C830 prep (BUCHI)

BUCHI pure C830 is a compact preparative instrument.

-Working pressure range : 0-300 bars

-Flow rate 0-100 mL/min, classically 6 or 20 mL/min in the lab

-columns : XBridge peptide BEH C18 OBD prep column (5 mm, 300 Å, 10 × 250 mm), XBridge Prep C18 5 µm OBD, 19 × 150 mm, or 19 × 100 mm and Zorbax 300SB-C3 column (5 µm, 9.4 × 250 mm) are placed in an extra temperature chamber (up to 70 °C).

– 2 pistons parallel arranges (4 different solvants) for binary gradient

-manual injection of liquid sample (20 mL loop)

-UV/Vis wavelength range (deuterium/halogen sources) : 200-800 nm

-diode array detection (DAD) – automated fraction collector

Pure C830 prep (BUCHI)